User's Guide

20 21
* Structure of DIABECARE DANA-i Menu:
3. Programming the Insulin Pump
Warning Follow the training and advice of a pump specialist Healthcare professional
and certified Insulin pump trainer whilst inputting the initial settings. Incorrect settings
may cause serious harm.
Warning
MAIN MENU
MAIN MENU
BOLUS BG BOLUS
CALCULATOR
MORE
INFORMATION
HISTORY SHIPPING
INFORMATION
BOLUS
SETTING
BASAL
REFILL OPTION
Back to
REVIEW
EDIT
BASAL
VIEW
BASAL
RATE
CHANGE
PROFILE
REFILL PRIME PRIME
CANNULA
CLOCK
SETTING
USER
OPTION
AIRPLANE
MODE
BOLUS SUSPENDREVIEW BUTTON
LOCK
TEMPORARY
BASAL
Follow up education is recommended for all insulin pump users.
1. When starting on insulin pump therapy, the patient should have daily contact with the
pump trainer and/or medical professional.
2. Visit with the Endocrinologist, Diabetologist or Advanced Practice Nurse within 3-7 days.
3. At first schedule weekly/biweekly consults then periodically as needed and advised.
4. Visit specialist monthly until pump regimen is established and then at least once every
three months or intervals advised by your medical professional.
DOCTOR MODE is a configuration menu accessed only by healthcare professionals and
certified insulin pump trainers. These settings are generally related to safety and to insulin
dosages about individual patients.
Contact your healthcare professional in order to change these settings.
Preset Bolus
Glucose Check alarm (min)
Bolus Block
Bolus Increment
Basal Increment
Ideal B.G level
Active Insulin
Insulin Decrese Ratio(%)
Maximum Basal (u/h)
Maximum Bolus (u)
Maximum Total daily does(u)
Safety Ratio (%)
Block Sensitive
Set UTC time (Date, Time)
4. Patient Education
* About Doctor Mode