User's Guide

22 23
1. Adjust the time
Setting the correct time is necessary for accurate basal insulin delivery and for retaining an
accurate record of all insulin delivery.
From within the Clock Setting menu adjust the time using
the or key. Press to save the setting.
Notice
Changing 12 or 24hour clock format refer to 3.4 Setting User Options.
Notice
The Diabecare DANA-i Insulin Pump has UTC time. Setting the date and time is
only completed within the Dr Mode.
OPTION
MAIN MENU
SUB MENU
CLOCK
SETTING
USER
OPTION
AIRPLANE
MODE
UTC = 0 Greenwich mean + 1 hour UK / Portugal / Europe West
-1 hour West Africa + 2 hour France / Germany / Italy
-2 hour Atlantic + 3 hour Europe East / Istanbul
-3 hour Atlantic + 4 hour Dubai
-4 hour US East + 5 hour Asia / Uzbekistan
-5 hour US Central / Chile + 6 hour India
-6 hour Canada + 7 hour Thailand
-7 hour US Pacific + 8 hour West Coast Australia / China
-8 hour Alaska + 9 hour Korea / Japan
-9 hour South Pacific Ocean +10 hour East Coast Australia
-10 hour Hawaii / Rarotonga + 11 hour Pacific / Noumea / Norfolk
-11 hour Samoa + 12 hour New Zealand
TIMEZONE : 00
UTC TIME
01/01/2019 00:00:00
LOCAL TIME
01/01/2019 00:00:00