User's Guide

30 31
1. TIME DISPLAY
Adjust the time display as 12hour or 24hour.
2. BUTTON SCROLL
When ON holding the or buttons adjusts the
value quickly.
3. BEEP
Key Beep ON/OFF enables an audio tone when
buttons are depressed.
4. ALARM
Change between SOUND, VIBRATION or BOTH for
alerts and pump alarms.
Notice
for safety some important alarms will
SOUND even though VIBRATION is selected.
5. LCD ON(S)
Adjust the duration the LCD remains on before
changing to Screen Saver Mode.
Set between (5 240) seconds.
6. BACKLIGHT ON(S)
Adjust the duration that the LCD backlight remains
on between button presses.
Adjust time for backlight from 0 sec to 60
7. LANGUAGE
Change different language option set by Country /
Region.
* User Options
USER OPTION
1.TIME DISPLAY:12
2.BUTTON SCROLL:ON
3.BEEP:ON
4.ALARM:SOUND
5.LCD ON(S):60
6.BACKLIGHT ON(S):10
7.LANGUAGE:EN
8.GLUCOSE UNIT:MG
9.SHUTDOWN:0
10.LOW RESERVOIR:20
11.PASSWORD
12.CANNULA VOL.:0.4
13.ADJ. RES. VOL.:245U
14.IDEAL BG:100
15.EXIT
8. GLUCOSE UNIT
Adjust the unit of measure for Glucose results
between ML (mmol/L) or MG (mg/dL).
Warning
Using wrong unit of measure could lead to
Glucose results being misinterpret.
9. SHUTDOWN
This is a safety setting, where if no buttons are
depressed after the time set (0 24) the pump stops
deliver and an alarm sounds.
Set the time to (0) to disable this auto off.
10. LOW RESERVOIR
Adjust the LOW RESERVOIR warning alarm threshold
(10, 20, 30, 40, 50) units of insulin remaining.
11. PASSWORD
Change the BUTTON LOCK password. Enter the
current PASSWORD and .
From the PASSWORD screen enter the new
password then to save. The password can be set
from 0 to 9 and A to F.
Notice
Default password is derived from the
manufacturing date and calculates as MMDD where
MM are the month and DD are the days , the pump
was producted. You can see the manufacturing date
in the Shipping Information menu, refer to the
chapter 6.6 Shipping information.
Caution
Password “0000” is easily unlocked.
This may be dangerous for children.
Notice
If forget your password number, contact
your dealer.
USER OPTION
1.TIME DISPLAY:12
2.BUTTON SCROLL:ON
3.BEEP:ON
4.ALARM:SOUND
5.LCD ON(S):60
6.BACKLIGHT ON(S):10
7.LANGUAGE:EN
8.GLUCOSE UNIT:MG
9.SHUTDOWN:0
10.LOW RESERVOIR:20
11.PASSWORD
12.CANNULA VOL.:0.4
13.ADJ. RES. VOL.:245U
14.IDEAL BG:100
15.EXIT
USER OPTION
10.LOW RESERVOIR:20
11.PASSWORD
12.CANNULA VOL.:0.4
USER OPTION
10.LOW RESERVOIR:20
11.PASSWORD:0000
12.CANNULA VOL.:0.4
PASSWORD
1 2 3 4
SAVE
SETTING?
:NO- OK :YES