12. CANNULA VOLUME:
Soft needle cannula has a pre-assigned volume
that needs to be filled with insulin before delivery.
Set the pre-set the required volume here for the
specific cannula used.
(Refer to 10.3 Prime Volume of infusion sets)
13. ADJ. RES. VOL:
This is where the remaining reservoir volume in the
pump can be adjusted.
Caution
Changing the volume to an incorrect
amount, may cause the pump to run out of insulin
without alarm.
14. IDEAL BG:
This is the ideal or Target BG value from the ‘BG
Bolus Calculatorwithin the pump bolus menu.
Use and key to adjust level .
15. EXIT:
Press to exit and save settings.
USER OPTION
14.IDEAL BG:100
15.EXIT
1.TIME DISPALY:12
* User Options
SAVE
SETTING?
:NO- OK :YES
USER OPTION
12.CANNULA VOL.:0.4
13.ADJ. RES. VOL.: 245U
14.IDEAL BG
USER OPTION
13.ADJ. RES. VOL.:245U
14.IDEAL BG:100
15.EXIT
USER OPTION
11.PASSWORD
12.CANNULA VOL.:0.4
13.ADJ. RES. VOL.:245U
Notice
When adjusting important USER OPTION settings like Language, Glucose Unit, or
Shutdown, a confirmation YES/NO is required.
5. More Information Screen
The More Information screen provides a quick review of:
Active Insulin from a previous bolus.
Extended bolus information (if active).
The most recent bolus delivery information including how many minutes ago the bolus
was delivery and the volume of the bolus.
1. From the main menu select REVIEW press .
2. Select MORE INFORMATION, press .
ACTIVE INSULIN
This is the Active Insulin sill working from previous
boluses.
DAILY TOTAL
Displayed in units for the current day.
EX T.B (Extended bolus)
If an Extended Bolus is active, the Bolus amount
and time remaining is displayed.
PRE. BOLUS (Previous bolus)
The most recent BOLUS is displayed as time since
bolus and bolus amount.
ACTIVE INSULINE: 4.2u
DAILY TOTAL: 25.0u
EXT. B:10.00u/00:30
PRE.BOLUS: 0h03m/3.5u
REVIEW
MORE INFORMATION
MORE
INFORMATION
HISTORY SHIPPING
INFORMATION
SUB MENU
MAIN MENU
REVIEW