4. Loading Insulin into the Pump
1. Preparation
Loading and refilling the Insulin Pump with insulin is a technical process which involves
medication (insulin) and sterile components.
It is recommended that:
Retrieve the insulin vial from the refrigerator and let it warm up to room temperature
before starting.
Place all necessary components on a clean dry surface with good lighting.
DANA Insulin Pump
Analog insulin (room temperature)
DANA Reservoir (3ml)
DANA Infusion Set
DANA Auto Setter
Linking screw
Alcohol swab (x 1)
Wash and dry hands before opening sterile packets and starting the refill process.
Follow advice and recommended guidance from the healthcare professional and insulin
pump trainer.
Warning Disconnect the insulin pump from the infusion set and body before opening
or starting any of the refill procedure. Insulin could be unintentionally delivered if the
pump is opened while still connected.
Warning
Notice
Do not reuse parts or all of an old infusion set or reservoir.
Notice
The room temperature in this manual is 15℃(59℉) ~ 30℃(86)