User's Guide

38 39
1. Loosen the linking screw until the shaft cap part
is covered by the head part (blue part). This is
important to adjust the length accurately.
2. Insert a reservoir filled with insulin
into the DANA Auto Setter and turn on
by pressing the button.
Caution
The end (hexagonal part) of the linking
screw should be engaged with the hexagonal hole
of DANA Auto Setter.
3. Press the button again to start measuring.
The reservoir will wind down and then count the
volume.
Notice
When the button is held and the motor is
operating you will notice the reservoir wind down
into the device.
4. The pump will automatically display correct refill
volume. Press OK to save and go to Refill-PRIME
Menu.
3. Adjust the length of linking screw with Auto Setter
DANA Auto Setter
Insert a battery into Auto Setter
The DANA Auto Setter is intended for adjusting the length of linking screw, measuring the
amount of insulin in the reservoir and sending it to the pump with the wireless
communication.
Open the battery compartment. Insert a battery
positive side up. Close battery compartment.
Notice
The DANA Auto Setter uses 1/2 AA size
3.6V battery.
The Auto Setter must be upright on a firm flat surface during usage.
Cover the reservoir cap (with small white plastic cap) when using Auto Setter to
prevent insulin leaking out.
Caution
* Using DANA Auto Setter :