User's Guide

40 41
4. Inserting the Reservoir into the Insulin Pump
1. Insert the reservoir with linking screw into the
Insulin Pump as shown.
TIP! When inserting the reservoir, rotate the
reservoir 90 degrees until the notch on the side of
the reservoir slide into place within the pump.
Gently let the reservoir and linking screw fall into
place.
2. Position the reservoir cap stright line with the
guide line, and turning clock-wise 45-degrees.
DO NOT push or force the reservoir into the Insulin Pump as this could damage the
Pump or force insulin from the reservoir.
Caution
Notice
If repeated attempts to insert the reservoir fail, use another new reservoir.
5. Connecting the Infusion Set to the Insulin Pump
Attach the Infusion Set Tube counter clockwise
into the reservoir compartment until it is firmly in
place.
TIP! The DANA insulin pump uses a proprietary
LH(Left Hand) lure connection between the insulin
pump and the Infusion Set tubing. Only DANA
Infusion Sets will connect to the DANA insulin
pump.
Notice
Hold the Insulin Pump upside down while
removing the white cap and connecting the tube
to avoid insulin leaking into the Insulin Pump.
Warning DO NOT use an Infusion Set if the package is damaged, inadvertently opened
or wet.
Warning
Guide line
Open Close