User's Guide

42 43
1. Move to REFILL, press .
2. Select REFILL menu, press .
3. You can adjust the insulin amount directly using
the and buttons.
4. To confirm the insulin amount press button.
TIP! Press and hold the and buttons to
change the value to quickly.
5. Press to save and go to Refill-PRIME Menu.
7. Prime the Infusion Set Tubing6. Rell
Prime every new Infusion set tube to displace air from within the tubing. Visually confirm
that all bubbles are primed from the Infusion Set tubing. Upon completion of refill process
confirm the basal is active and correct.
With the refill input, pump get to know exactly what the insulin amount is.
Warning It is important to properly prime the Tube and ensure all air is removed from
the system. The pump may not properly deliver insulin without this.
Caution PRIME is a very important process to ensure that the pump will deliver insulin
accurately. Delivery problems often result due to air within the tube and occlusion
alarms may be because of poor or insufficient PRIME. Patients are required to have
good level understanding of how to properly PRIME and why the PRIME process is
important.
Warning
Caution
Notice
Connect the infusion set tube then position the pump upright during priming for
the perfect removal of any air in the reservoir and tube.
Remained air
Remained air
REFILL PRIME PRIME
CANNULA
REFILL
SUB MENU
MAIN MENU
CONFIRM
REFILL 280U?
:NO- OK :YES
REFILL
REFILL
MAIN MENU
REFILL
REFILL
0 u
REFILL
280 u