User's Guide

44 45
6. The PRIMING display will show the volume of
insulin delivered.
Notice
During PRIME the pump may BEEP or
VIBRATE after every unit of Insulin is primed.
7. When droplets of Insulin appear at the end of the
TUBE press to PAUSE. Check the entire length
of TUBE for any bubbles. Press to finish the
PRIME.
Warning
Ensure droplets of Insulin are clearly
visible at the end of the tube / needle before
stopping the prime
Notice
Unless stopped already the PRIME will
automatically stop after 25 units.
Notice
If the prime amount is not enough (less
than 7U), this message will be displayed. Because,
the minimum prime amount of the infusion set
connected to the DANA pump is 7U. Refer to 10.3
Prime Volume of infusion sets and Chapter 7. Alarms
and Error messages.
ALERT
PRIME AMOUNT
IS NOT ENOUGH
* PRIME procedure:
1. From the main menu select REFILL and press .
2. Select PRIME from the REFILL's sub menu and
press .
3. From the PRIME menu press to start.
Warning
It is very dangerous to start a Prime
whilst the Infusion set is connected to the body.
4. Stand PUMP upright during PRIME, press .
Air moves upwards to top–standing pump
upright helps displace all air bubbles.
5. START TUBE PRIME confirmation menu, press
to start PRIME
MAIN MENU
REFILL
REFILL
PRIME
:EXIT- OK :START
PRIME
0.0 u
INSERT RESERVOIR/
CONNECT INFUSION/
UPRIGHT PUMP
DURING PRIME
START
TUBE PRIME?
:NO- OK :YES
REFILL PRIME
CANNULA
SUB MENU
MAIN MENU
REFILL
PRIME
:EXIT- OK :RE-START
PAUSE
12.5 u