User's Guide

46 47
5. Press to confirm.
6. The PRIME CANNULA window displays the
delivery.
Notice
Cannula fill volume is set in the USER OPTION menu. Cannula fill can be set
between 0.1 – 0.9 units. Read cannula instructions to determine individual requirements
for filling.
START
CANNULA PRIME?
:NO- OK :YES
PRIME CANNULA
0.1 u
8. Prime the Cannula
When using an infusion set with a soft needle/cannula, the hollow area within the cannula
requires PRIME CANNULA after completing tube prime.
1. After inserting the cannula into the body
(following the cannula instructions) connect the
infusion set tube to the infusion set after priming
all bubbles out from the tube.
2. Select REFILL menu from the MAIN MENU and
press
.
3. Select PRIME CANNULA from the REFILL
menu press .
4. Check the PRIME CANNULA VOLUME is
correct press to confirm.
REFILL
PRIME CANNULA
MAIN MENU
REFILL
PRIME CANNULA
:EXIT- OK :START
0.4 u
REFILL PRIME
SUB MENU
MAIN MENU
REFILL
PRIME
CANNULA