User's Guide

80 81
When battery energy is low, an empty battery
symbol appears and blinks on the initial screen.
BT Communication is interrupted to conserve the
battery.
Alarm will be activated with sound and repeat every
10 minutes or whenever the pump is awoken from
the screen save mode until replace the battery.
* ALARM Message
Low Battery warning
LOW BATTERY
The low battery screen will appear when the battery level is not sufficient to operate
the pump and deliver insulin.
How to solve:
Silence the alarm by pressing any button and replace the new AAA battery as soon as
possible.
03/10/2019
u/h 100%
245u
0.20
10:04 AM
B
Caution Always ensure you keep spare batteries near.
Caution
ALARM
LOW
BATTERY
7. Alarm, Error and Alert
This chapter describes insulin pump Alarm, Error, Alert messages and how to solve them.
Type Message
ALARM
(High Priority)
LOW BATTERY
EMPTY RESERVOIR
SHUTDOWN
OCCLUSION
CHECK ERROR
SYSTEM ERROR
PRIME INCOMPLETE
BASAL MAX
DAILY MAX
ERROR
(Medium Priority)
LOW RESERVOIR
SUSPEND
DELIVERY LESS THAN BASAL SET RATE
ALERT
(Low Priority)
MISSED BOLUS
PRIME AMOUNT IS NOT ENOUGH
CHECK GLUCOSE
CONFIRM PAIRING
NO DELIVERY
Notice
Alarm is a critical alarm that can affect safety. Resolve an issue as soon as possible. In this
case, the alarms will SOUND even though VIBRATION is selected.
Error makes you know the problem of the insulin pump. An ERROR is less serious than
an ALARM.
Alert just informs you about the status of the insulin pump or if you need to make a
decision.