User's Guide

82 83
OCCLUSION
This alarm occurs if the Insulin Pump has an occlusion or
a problem which disturbs insulin delivery.
How to solve:
An occlusion alarm will occur when the Insulin Pump detects a blockage and cannot deliver
insulin. Check for blocked or folded areas and replace the reservoir or infusion set if necessary.
Caution
Even after resolution of the problem - check your blood glucose frequently
to ensure the pump is delivering insulin properly.
ALARM
OCCLUSION
Self-check Procedures for Occlusion Alarm Occurrence
Implement self-check procedures in the case of the following:
An occlusion alarm occurs during replacement of the infusion set or reservoir.
The occlusion alarm occurs frequently.
Step.1 Safety first - check BG levels (could be Hyperglycemia)
Step.2 Visually check if there is any area of the tubing that is folded or blocked.
Step.3 To determine of the occlusion is in the pump or body/consumable:
a. Disconnect infusion set from the body.
b. Deliver a BOLUS of 5 - 6 units.
c. If there is no occlusion alarm or blockage, it will be possible to visually notice/
see a puddle of insulin at the end of the Infusion Set tubing. This has now
determined that the occlusion was in the cannula or body. Replace cannula
or insertion site to resolve.
* ALARM Message
SHUTDOWN
The Pump will automatically give an alarm sound if no
buttons are pressed after the pre-set shutdown period
is exceeded. If no acknowledgment of the alarm is
received (button press) following the audible alarm -
the pump will suspend all insulin delivery.
EMPTY RESERVOIR
When the reservoir volume is zero (0u), all delivery is
stopped and this screen will be shown with a sound
alarm.
How to solve:
Silence by acknowledging the alarm and pressing any of the buttons.
How to solve:
Silence the alarm by pressing any button. Immediately replace the reservoir and refill
the pump
Warning
The pump displaying "NO DELIVERY/EMPTY RESERVOIR" is unable to not
only deliver basal and bolus but access to any delivery function.
Notice
This alarm will repeat every 5 minutes until a complete refill is completed. Refer
to chapter 4. Loading Insulin into the Pump.
ALARM
SHUTDOWN
ALARM
EMPTY
RESERVOIR