User's Guide

86 87
DELIVERY LESS THAN BASAL SET RATE
If the basal is skipped and delivered less than 80% of
the basal setting, this alarm will be generated.
How to solve:
Silence the alarm by pressing any button. If you operate the pump at basal delivery
interval, the basal may occasionally skip. For a stable basal delivery, avoid long-time
pump button operation.
Basal Insulin delivery intervals very based on the size of the set basal rates.
Size of BASAL Rate (u/hr) BASAL delivery interval
0.1 U/h (Basal)
Every 4 minutes (1/15) of the hourly rate is delivered.
15 deliveries per hour.
Extended bolus
0.09 U/h (Basal)
Basal delivery will occur once at 56min the hour.
(hourly)
Warning
The individual small basal delivery increments maybe interrupted during
Bluetooth pairing or during changes being made to configuration or pump settings.
These increments of basal delivery in very low basal delivery rates such as ≤ 0.09 U/h
patients need to be monitored carefully to avoid unexpected hyperglycemia which it
could lead to ketoacidosis
ERROR
DELIVERY LESS THAN
BASAL SET RATE
* ERROR Message
LOW RESERVOIR
When the reservoir volume is below the 'Low Reservoir'
warning configured in the user options, this screen will
be shown with either beep or vibrates.
SUSPEND
Select any menu related to infusion (insulin delivery)
whilst the Insulin Pump is in Suspend Mode, you are
alerted with this message.
Refer to 6.2 SUSPEND.
How to solve:
The pump will revert to the Initial Display and the reservoir icon will blink/flash. After
checking the actual remaining insulin volume of the reservoir in your pump, replace the
reservoir and refill the pump if necessary.
How to solve:
Turn the Suspend Mode off prior to continuing in any of the infusion (insulin delivery)
menu's.
Notice
The pump has a Low Reservoir alert. Depending upon dosage-the user can set
this alert at 10, 20, 30, 40 or 50u threshold (See section 3.4 User option for changing the
threshold). If set at 10 or 20u the pump will alert every half from when the threshold is
reached. If set at 30, 40 or 50u the pump will alert very 1 hour after the threshold is
reached.
ERROR
LOW
RESERVOIR
ERROR
SUSPEND