CONFIRM PAIRING
Displayed when the pump receives a pairing signal.
CHECK GLUCOSE
This alarm is to remind you to check blood glucose level
after a bolus.
How to solve:
Silence the melody by depressing any button and check your blood glucose level.
How to solve:
To cancel / prevent pairing, press NO.
Refer to Manual of Mobile application.
Notice
The default is 2 hours and can be adjusted by a health care professional in the
Dr. Mode.
CONFIRM
PAIRING?
:NO- OK :YES
ALERT
CHECK
GLUCOSE
* ALERT Message
PRIME AMOUNT IS NOT ENOUGH
This alert message will be displayed if the volume
delivered for tubing prime is less than 7 units.
How to solve:
Properly priming the infusion set tube is necessary to ensure all air is displaced and
insulin is ready for infusion. Even the shortest infusion set tube will require more than 7
units to properly prime - so for safety the Insulin Pump has a minimum required Prime
volume of 7 units. Refer to 10.3 Prime Volume of infusion sets for suggested minimum
prime amount for each infusion set.
ALERT
PRIME AMOUNT
IS NOT ENOUGH
MISSED BOLUS
If a bolus was missed during the time period you had
set, the Insulin Pump will give you an alert together with
an message. Refer to 5.3 Bolus Setting-MISSED BOLUS
How to solve:
Silence the alert by pressing any button.
Follow the prompt by determining if a food bolus was missed and administer if
necessary.
ALERT
MISSED BOLUS
01:00PM - 02:00PM