User's Guide

90 91
This page left blank intentionally
Reason of NO DELIVERY
Pump has been suspended. Refer to 6.2 suspend
If there is no insulin in the reservoir, EMPTY
RESERVOIR is displayed and insulin is not injected.
Refer to 4. Loading Insulin into the Pump.
If the warning of BASAL MAX, DAILY MAX is
activated, the delivery could be restricted for an
hour/day
03/10/2019
0u
10:04 AM
EMPTY RESER.
03/10/2019
0u
10:04 AM
SUSPEND
If prime is not completed fill tube. Refer to 4.7
Prime the infusion set tubing.
If basal setting is 0.0 u/h, NO DELIVERY is displayed
during that time.
Refer to 3.2 Setting the Basal rate.
03/10/2019
245u
10:04 AM
NO PRIME
03/10/2019
245u
10:04 AM
DAILY MAX
03/10/2019
u/h 100%
245u
0.00
10:04 AM
NO DELIVERY
The pump cannot deliver insulin for one or more of
different reasons. This message is shown on initial
display and may blink/flash with additional information
How to solve:
A detail message will blink alternately. Refer to following reason of NO DELIVERY.
03/10/2019
245u
10:04 AM
NO DELIVERY
* ALERT Message