User's Guide

8 9
1. Pump users need more than 4 blood glucose measurements per day, and vision and
hearing to receive any pump alarm.
2. Patients must not open the Pump housing or handle any internal components.
3. The Diabecare DANA-i Insulin Pump is intended for use with a proprietary Infusion Set,
reservoir and other accessories specified in this booklet. DO NOT use the Pump with any
other infusion system or accessories.
4. Press buttons with the pad of the finger. DO NOT use fingernails or any sharp objects.
5. The Insulin Pump comes with factory default settings and alarms, maximum daily totals,
basal and bolus doses. These settings can be adjusted by a healthcare professional.
6. The pump is indicated for use with U-100 NovoLog insulin. The other insulin has not
been tested and may not be compatible for use with the Diabecare DANA-i insulin
pump.
7. Change the reservoir and the Infusion Set regularly, as recommended by your healthcare
professional. DO NOT use for longer than the intended period.
8. Check the expiration dates and dispose of any expired accessories.
6. Precautions
Glucose Check Alarm 0min
Maximum Daily Total 80u
Maximum Bolus 40u
Maximum Basal 3.3u/h
9. Avoid impact damage such as dropping. If there is any known damage of pump and
accessory, contact your healthcare professional or technical support from the local
Insulin Pump distributor.
10. For any trouble with any of the system components, turn off the Insulin Pump by
removing the battery and contact a healthcare professional or Insulin Pump trainer.
11. Remove the battery for long-term storage.