User's Guide

14 15
Warning Change the battery in a clean dry environment to prevent water/ ingress from
entering the pump case. The battery cap is correctly installed with a push and turn
counter clock-wise to locked. This prevents water/ingress.
Warning
Notice
The Diabecare DANA-i Insulin Pump is powered by an external (AAA) Battery
Notice
SOOIL recommend using either a Duracell gold alkaline AAA battery or an Energizer
Advanced’ alkaline AAA battery.
Caution The pump required one AAA 1.5V battery. Use a new AAA alkaline battery. Do
not use a carbon zinc battery in your pump. Carbon zinc batteries are not compatible
with this pump.
Caution Lithium batteries are not recommended as the battery level indicator may not
be accurate.
Caution
DO NOT attempt to change the battery while a bolus is in progress.
Dispose of used batteries in an environmentally friendly way according to local
disposal requirements or contact your local insulin pump distributor for disposal
information.
It is recommended to keep a spare battery as backup.
For accurate reading of the remaining battery charge, check the battery display
following the delivery of a bolus
Caution
Caution
Caution
1. Open the battery cap and turning
clock-wise 45 degrees.
2. Insert the battery with the positive (+) at
top and insert the negative (-) into the
Insulin Pump.
3. Replace the battery cap turning it
counter clock-wise.
4. Completed when the cap is firmly
locked with insulin pump.
2. Installing a battery
Battery Cap
With O-ring
A
A
A
Open Lock