Instruction Manual
6/8
Form 948 (10.09) ©2009 SOR Inc. 
Typical Fixed Tank Installations
GENERAL NOTES:
1.  13mm (0.52”) mounting hole required on right-angle brackets.
2.  Secure cable to bulkhead or tank wall using cable clips for 7mm - 8mm (0.28” - 0.32”) 
  dia. cable. Maximum distance between clips is 500mm (19.69”). Avoid strain on cable. 
3.  Use bulkhead seals suitable for 7mm - 8mm (0.28” - 0.32”) cable. 
4.  Terminate vent tube and connect cable within an environmentally protected enclosure.
15.0 
MIN
0.59
DO NOT use 
transmitter 
as a step!
Support cable 
on bulkhead 
or cable tray.
Sludge Level Maximum
15.0 
MIN
0.59
249.9 
MIN
 9.84
Sludge Level Maximum
Linear = mm/inches
Sludge Level Max.
 300 
MIN
11.81
35.0 
MIN
1.38