User Manual

Table Of Contents
12. IMPLANT PROCEDURE
NECESSARY EQUIPMENT
Implantation of ALIZEA DR / SR requires the following equipment:
MicroPort dedicated programmer, equipped with the SMARTVIEW software interface
and inductive telemetry head,
pacing system analyzer, as well as its sterile connecting cables, to evaluate the pacing
and sensing thresholds,
a complete set of leads with corresponding introducers,
physiological signal monitor capable of displaying simultaneously the surface ECG and
arterial pressure,
an external defibrillator with sterile external paddles,
sterile cover for the telemetry head.
PACKAGING CONTENTS
ALIZEA DR / SR and its accessories are ethylene oxide sterilized and hermetically sealed in
two-ply clear packaging meeting international requirements.
The sterile packaging contains:
the pacemaker,
a ratcheting screwdriver.
As delivered, ALIZEA DR / SR is programmed to as-shipped values that are different from
nominal values (see “Programmable Parameters” section for details).
OPTIONAL EQUIPMENT
The following equipment may be required during implantation of ALIZEA DR / SR:
an IS-1 insulating plug to close the atrial port (only for ALIZEA DR),
sterile water to clean traces of blood. Any parts cleaned with sterile water must be
thoroughly dried,
mineral oil to lubricate if necessary,
a lead cap to isolate a lead which is not used.
OPENING THE PACKAGE
Before opening the package:
Check the "Use by" date printed on the labels on the box and on the sterile package.
Pacemakers not implanted before the “Use by” date should be returned to your MicroPort
representative.
Interrogate the device before activating the remote monitoring: if a warning is displayed
or if battery voltage is too low (< 3.04V), do not implant the device and contact your
MicroPort representative.
12.1.
12.2.
12.3.
12.4.
12. IMPLANT PROCEDURE
30 ALIZEA – UA10414A