User Manual

Table Of Contents
CHOOSING THE TYPE OF LEAD
For the DR model, for optimal use of pacemaker functions, the atrial lead should be bipolar.
If a unipolar atrial lead is used, the Autosensing function will not be available. For MRI
compatibility, both leads should be bipolar.
For the SR model, for optimal use of pacemaker functions, the lead used should be bipolar. If
a unipolar lead is used, the following function will not be available: the rate-response function
based on minute ventilation. For MRI compatibility, the lead should be bipolar.
Use of a high polarization ventricular lead interferes with normal operation of the Ventricular
Autothreshold function.
Connections:
The pacing/sensing connectors are IS-1 compatible. Other lead configurations are not
compatible.
MEASUREMENT OF THRESHOLDS AT IMPLANT
Pacing and sensing thresholds should be measured at implant.
Pacing thresholds:
Acute thresholds should be lower than 1 V for a 0.5 ms pulse width, both in the ventricle and
in the atrium.
Chronic thresholds should be less than 3 V for a 0.5 ms pulse width.
Sensing thresholds:
For proper ventricular sensing, the amplitude of the R-wave should be greater than 5 mV. For
proper atrial sensing, the amplitude of the P-wave should be greater than 2 mV.
LEAD CONNECTIONS
It is imperative that each lead be properly connected to the corresponding pacemaker
connector (the position of each connector is indicated on the casing).
CAUTION:
For the SR model, only one lead will be connected (ventricular or atrial). For the DR model,
two leads will be connected: ventricular and atrial.
For each lead, only distal inserts must be tightened. To connect each lead, proceed as
follows:
1. For the SR model, insert the ventricular or atrial lead pin all the way into the pacemaker
port.
For the DR model, insert the ventricular lead pin all the way into the appropriate
pacemaker port (marked V) of the pacemaker.
2. Insert the screwdriver in the centre of the prominent screwdriver slot contour on the distal
insert, then into the pre-inserted screw.
3. Tighten the screw until the screwdriver clicks. Check the fixation by pulling the lead (the
lead should remain in the pacemaker when pulling to check the tightness).
12.6.
12.7.
12.8.
12. IMPLANT PROCEDURE
32 ALIZEA – UA10414A