4. For the DR model, insert the atrial lead pin all the way into the appropriate port (marked
A) of the pacemaker and proceed as in steps 2 and 3.
CAUTION:
Do not tighten the pre-inserted screw when there is no lead connector inserted (this could
damage the connector block).
Do not loosen the screw before inserting the lead into the connector (subsequent risk of
being unable to reinsert the screw).
Use only the screwdriver provided with the pacemaker. Maintain the screwdriver's shaft
perpendicular to the plane of the pacemaker (see figure below).
WARNING: Ensure that the screwdriver's tip is fully inserted in the setscrew; otherwise, the
screwdriver might damage the setscrew and prevent connection with or disconnection from
the lead.
To ensure full insertion, push the screwdriver's hex tip smoothly into the setscrew until it
reaches the bottom of the hex chamber in the screw, which can be felt as a solid metallic
contact. Do not implant the pacemaker if there is no feeling of solid metallic contact. Do
not implant the pacemaker if the screwdriver does not click when attempting to tighten the
setscrew on the lead pin.
IMPLANT CONSIDERATIONS
After placing the device in its pocket, check the pacemaker functioning using an ECG.
Suture the device to the muscle in order to avoid potential migration of the device into the
pectoral muscle.