THERAPEUTIC IMPROVEMENTS
Features DR model SR model
SafeR x
Dplus x
Rate acceleration x
Rest rate x x
Smoothing x x
Mode switching x
SafeR mode
The SafeR mode is intended to minimize unnecessary ventricular pacing and to promote
spontaneous atrioventricular conduction. It gives priority to pacemaker operation in AAI(R)
mode: the sensed P wave does not trigger the AV delay and recycles an atrial escape
interval.
Indication:
SafeR is designed for patients requiring AAI(R) pacing (sinus node dysfunction) and/or who
are at risk for AV conduction disorders (paroxysmal AV block, permanent first degree AV
block, or exercise-induced AV block).
Operation:
The pacemaker functions in AAI(R) mode and temporarily converts to DDD(R) mode in the
following cases:
AVB III: Two consecutive atrial events (paced or sensed) occur without subsequent
ventricular beats.
AVB II: Three atrial events in twelve consecutive atrial cycles without subsequent
ventricular beats.
AVB I: Six consecutive atrial events occur without subsequent ventricular beats within
a programmed PR interval (Long PR min and Long PR max parameters). This AVB I
criterion can be activated continuously or only during exercise phase depending on the
AVB I switch parameter.
PAUSE: No ventricular beat occurs within a programmed interval (Max pause parameter,
programmable to 2, 3 or 4 seconds).
The device regularly switches back to AAI(R) mode (at least once a day):
after sensing 12 consecutive spontaneous ventricular events
automatically every 100 paced ventricular cycles.
After an automatic switch to AAI(R) mode, if the intrinsic AV conduction has not resumed,
the device switches back to DDD(R) mode according to the criteria listed above (third degree
AVB, second degree AVB, first degree AVB or the pause).
The SafeR-R mode is a SafeR mode with Rate Response.
The SafeR/DDIR mode activates a switch from SafeR to DDIR mode when the pacemaker
enters Fallback Mode Switching. Rate responsive pacing is only functional when the patient
is in Fallback Mode Switching. The physical activity used during the Fallback Mode Switching
phase is equal to the programmed parameter.
15.6.
15.6.1.
15. FUNCTIONS AND PARAMETERS
ALIZEA – UA10414A 43