High pacing threshold (A,V);
Lead Polarity Switch (A,V);
AF occurrence with 3 thresholds (for DR only);
Fast V rate during AF (for DR only);
Asynchronous mode;
MRI notifications.
The following set of alert triggers (system alerts) cannot be deactivated when the Alerts are
programmed “On” and can trigger an alert transmission:
Battery depletion – RRT;
System integrity;
Reset.
WARNING: The use of remote monitoring does not replace regular follow-up. Therefore,
when using remote monitoring, the time period between follow-up visits should not be
extended.
The pacemaker uses BLE technology with authenticated pairing with the Smartview Connect
monitor, and data encryption for secure communication.
An excessive BLE usage is an indicator for a possible cyberattack. In this case, the BLE
communication may be automatically forced to off. In such event, a warning will appear on
the programmer during a follow-up visit to acknowledge the physician.