User Manual

Table Of Contents
16. PATIENT FOLLOW-UP
RECOMMENDED REPLACEMENT TIME (RRT)
The Recommended Replacement Time (RRT), previously known as Elective Replacement
Indicator (ERI) is controlled by:
a battery voltage:
of 2.63 V if Remote monitoring activated;
of 2.70 V if Remote monitoring deactivated.
When RRT is reached, a magnet rate of 80 ± 1 ppm is applied by the device. As the EOS
(End Of Service) approaches, the magnet rate decreases to 70 ± 1 ppm.
CAUTION:
The pacemaker should be replaced as soon as the Recommended Replacement Time (RRT
or ERI) point is reached.
When the RRT (or ERI) point is reached, the following parameters are switched to the settings
specified below until the end of the device's service:
Mode: VVI
Basic rate: 70 ppm
Rate hysteresis : off
Rate response mode: off
Smoothing: off
Other parameters remain as programmed.
If activated, the remote monitoring function is deactivated after RRT once the RRT alerts have
been sent.
After having reached the RRT (or ERI), the Prolonged Service Period (PSP) of the device is:
at least 3 months under the following conditions: VVI, 70 ppm, 5V, 0.35ms, 500 Ω, 100%
pacing (Nominal parameters).
at least 6 months under the following conditions: VVI, 60 ppm, 2.5V, 0.35ms, 600 Ω,
100% pacing according to ISO14708 section 19.2.
Time to RRT (residual longevity): estimated remaining service time of the device. This
estimate is calculated using battery voltage, leads measurements, statistics and programmed
parameters. This value is refreshed automatically after each programming impacting device
longevity.
FOLLOW-UP RECOMMENDATIONS
Standard follow-up
Annual standard follow up is recommended. Before the patient’s discharge, and at each
follow-up visit, it is advisable to:
1. Check the battery status.
16.1.
16.2.
16. PATIENT FOLLOW-UP
54 ALIZEA – UA10414A