User Manual

Table Of Contents
2. Check for proper sensing (sensitivity for the both SR and DR models, crosstalk
for the DR model) and pacing; set the pacing amplitude to twice the pacing threshold.
3. Interrogate the device memories (AIDA)
4. Keep a printout of programmed parameters, test results, and memory data.
Frequent follow-up
When the minimum estimated residual longevity displayed by the MicroPort dedicated
programmer is less than or equal to 12 months:
It is recommended to conduct patient follow-up visit at an interval that is between
the minimal and the typical residual longevities displayed by the MicroPort dedicated
programmer, without exceeding 12 months (i.e. the annual standard follow-up).
For pacemaker-dependent patients, it is recommended to conduct patient follow-up
visit at an interval equal to the minimal residual longevity displayed by the MicroPort
dedicated programmer.
NOTES:
“Time to RRT = N/A” is displayed by the MicroPort dedicated programmer
If less than 5 minutes statistics are available (i.e. first interrogation at implant, device
reset),
If any lead impedance related to each currently pacing chamber does not have a valid
value (range between 200 and 3000 Ω).
If “Time to RRT” is still N/A (outside of the two cases listed here above), contact your
MicroPort Representative.
Replacement
The pacemaker should be replaced as soon as the Recommended Replacement Time (RRT)
point is reached.
Implant software upgrade
In case new implant software is downloaded in the device memory through the MicroPort
dedicated programmer, a warning message may be displayed by the MicroPort dedicated
programmer to inform the user and give the proper instructions to follow.
EXPLANTATION
The pacemaker should be explanted in the following cases:
End of service;
Confirmed malfunction;
Cremation of the patient (the pacemaker may explode if placed in an incinerator);
Patient death: for environmental reasons, the local regulation may require the
explantation of the devices containing a battery supply.
WARNING:
The pacemaker is a single use impantable device. The explanted pacemaker should not be
reused in another patient. It could lead to inflammation, intoxication or infection of the patient.
16.3.
16. PATIENT FOLLOW-UP
ALIZEA – UA10414A 55