User Manual

Table Of Contents
Clinical alerts Models Values Nominal
value
"As
shipped"
value
AT/AF burden DR
ON-OFF OFF OFF
Low AF burden
daily limit (h)
DR
6 min-15 min-30 min-1h-3h-6h-
12h-24h
6 min 6 min
Mid AF burden
daily limit (h)
DR
OFF-6 min-15 min-30 min-1h-3h-
6h-12h-24h
6h 6h
High AF burden
daily limit (h)
DR
OFF-6 min-15 min-30 min-1h-3h-
6h-12h-24h
24h 24h
Fast V Rate during
AT/AF
DR
ON-OFF OFF OFF
Fast V Rate limit
(ppm)
DR
80-90-100-110-120 100 100
Fast V Duration
limit (h)
DR
0.5-1-3-6-12-24 1 1
Other alerts Models Values Nominal
value
"As
shipped"
value
MRI notifications SR, DR
ON-OFF OFF OFF
Asynchronous
mode
SR, DR
ON-OFF ON ON
21. PROGRAMMABLE PARAMETERS
74 ALIZEA – UA10414A