21. PROGRAMMABLE PARAMETERS
Pacing/Sensing Values
Atrial sensitivity (mV)
(1)
From 0.2 to 4 by steps of 0.2 ; 0.4 (± 50 %)
Atrial amplitude (V)
(2)
1-1.5-2-2.5-3-3.5-4-4.5-5-6 (± 20 %)
Atrial pulse width (ms) 0.12-0.25-0.35-0.5-0.6-0.75-0.85-1 (± 10 %)
Ventricular sensitivity (mV)
(1)
From 0.4 to 4 by steps of 0.2 ; 0.4 (± 50 %)
RV amplitude (V)
(2)
1-1.5-2-2.5-3-3.5-4-4.5-5-6 (± 20 %)
RV pulse width (ms) 0.12-0.25-0.35-0.5-0.6-0.75-0.85-1 (± 10 %)
LV amplitude (V)
(2)
0.10-0.25-0.35 (± 50 %)
0.50-0.60-0.75-0.85 (± 30 %)
1.00-1.10-1.25-1.35-1.50-1.60-1.75-1.85-2.00-2.10-2.25-
2.35-2.50-2.60-2.75-2.85-3.00-3.10-3.25-3.35-3.50-3.60-
3.75-3.85-4.00-4.10-4.25-4.35-4.50-4.60-4.75-4.85-5.00-
5.10-5.25-5.35-5.50-5.60-5.75-5.85-6.00-6.10-6.25-6.35-
6.50-6.60-6.75-6.85-7.00- (± 20 %)
LV pulse width (ms) 0.12-0.25-0.35-0.50-0.60-0.75-0.85-1.00-1.10-1.25-1.35-
1.50-1.60-1.75-1.85-2.00 (± 10 %)
LV pacing polarity LV bipolar-LV tip to RV ring-LV ring to RV coil-LV tip to RV
coil-LV ring to CAN-LV tip to CAN
V chambers Right-R+L-L+R
VV delay (ms) 0-16-24-32-40-48-56-64- (± 3 ms)
(1) Values are measured using a positive and negative triangular signal of 2/13 ms.
(2) The correlation between the programmed amplitudes, the stored amplitudes and the mid-pulse
delivered amplitudes under a 500 ohm load are given in the following table:
Programmed amplitude (V) Stored amplitude (V) Mid-pulse delivered
amplitude (V)
0,10* 0,20 0,19
0,25* 0,26 0,22
0,35* 0,39 0,33
0,50* 0,52 0,44
0,60* 0,72 0,61
0,75* 0,85 0,72
0,85* 0,98 0,83
1,00 1,11 0,94
1,10* 1,24 1,05
1,25* 1,37 1,16
1,35* 1,50 1,27
1,50 1,63 1,38
1,60* 1,76 1,49
56
SORIN PLATINIUM CRT-D U459A