User Manual
Table Of Contents
- 1. General description
- 2. Indications
- 3. Contraindications
- 4. Warnings and precautions
- 5. Adverse events
- 6. Clinical studies
- 7. Patient selection and treatment
- 8. Patient counselling information
- 9. Conformance to standards
- 10. Physician guidelines
- 11. Patient information
- 12. How supplied
- 13. Device description
- 14. Implant procedure
- 14.1. Necessary equipment
- 14.2. Packaging
- 14.3. Optional equipment
- 14.4. Before opening the package
- 14.5. Prior to implantation
- 14.6. Device placement
- 14.7. Choosing the type of lead
- 14.8. Shock configuration (+ -> -)
- 14.9. Measurement of thresholds at implant
- 14.10. Leads connection
- 14.11. Device implantation
- 14.12. Tests and programming
- 15. Special modes
- 16. Main functions
- 17. Patient follow-up
- 18. Supplemental Information
- 19. Physical characteristics
- 20. Electrical characteristics
- 21. Programmable parameters
- 22. Non programmable parameters
- 23. Limited warranty
- 24. Patents
- 25. Explanation of symbols
21. PROGRAMMABLE PARAMETERS
21. PROGRAMMABLE PARAMETERS
measured at 37 °C under a 500 ohm load
Legend:
Value in bold: "as shipped" value
Underlined value: nominal value
21.1. ANTIBRADYCARDIA PACING
Basic parameters Values
Mode VVI-VVIR-DDD-DDDR-DDD/DDIR-DDI-DDIR-SafeR (AAI
<=> DDD)-SafeR-R (AAIR <=> DDDR)
Basic rate (bpm)
(1)
From 30 to 90 by steps of 5 ; 60 (± 4 %)
Maximum rate (bpm) From 100 to 145 by steps of 5 ; 120 (± 6 %)
Rate hysteresis (%) 0-5-10-20-35 (± 18 ms)
Rest AV delay (ms) 30-40-45-55-65-70-80-85-95-100-110-115-125-135-140-150-
155-165-170-180-190-195-205-210-220-225-235-250
(± 19 ms)
Exercise AV delay (ms) 30-40-45-55-65-70-80-85-95-100-110-115-125-135-140-150-
155-165-170-180-190-195-205-210-220-225-235-250
(± 19 ms)
AVD Paced/Sensed Offset (ms) 0-10-15-25-30-40-45-55-65-70-80-85-95-100-110-115-125
(± 1 ms)
(1) The corresponding periods are (in ms): 2000-1714-1500-1333-1200-1091-1000-923-857-800-750-
706-667 ms.
Special features Values
Smoothing OFF-Very slow-Slow-Medium-Fast
Mode Switch ON-OFF
Physical activity Very low-Low-Medium-High-Very high
52