21. PROGRAMMABLE PARAMETERS
Detection criteria Values
Slow VT and VT detection criteria Rate Only-Stability-Stability+-Stability/Acc-Stability+/Acc-
PARAD-PARAD+
Fast VT detection criteria Rate+Stability-Rate Only
Majority: (X/Y), Y (cycles) 8-12-16
Majority: (X/Y), X (%) 65-70-75-80-90-95-100
Window of RR stability for Slow VT
and VT (ms)
30-45-65-80-95-110-125
Window of RR stability for fast VT
(ms)
30-45-65
Acceleration (%) 6-13-19-25-31-38-44-50
Long cycle persistence extension
(cycles)
From 0 to 16 by steps of 1 ; 10
Long cycle gap (ms) 15-30-45-65-80-95-110-125-140-155-170-190-205
Atrial monitoring Yes-No
21.3. VENTRICULAR TACHYARRHYTHMIA THERAPIES
Common parameters Values
Enable ATP therapy Yes-No
Enable shock therapy Yes-No
ATP pacing chamber Right-Left-R+L
Polarity alternation (42J) Yes-No
Atrial coil (SVC) present Yes-No
Active case Yes-No
Shock configuration (+ --> -) Case to RV-SVC to RV-Case + SVC to RV-RV to Case-RV to
SVC-RV to Case + SVC
SVC exclusion (shock < 15J) Yes-No
SORIN PLATINIUM CRT-D U459A