User's Manual

Table Of Contents
22. NON PROGRAMMABLE PARAMETERS
22. NON PROGRAMMABLE PARAMETERS
Interval Values
Committed period 95 ms (± 5 ms)
Atrial refractory periods Values
Post atrial sensing 47 ms (± 16 ms)
Post atrial pacing 109 ms (± 4 ms)
Ventricular refractory periods Values
Post ventricular sensing 95 ms (± 16 ms)
Post ventricular pacing 220 ms (± 4 ms)
Post atrial pacing (blanking) 16 ms (± 3 ms)
Tachycardia criteria Values
Window of PR association 63 ms (± 1 ms)
Therapies Values
Waveform
(1)
Constant tilt (50% - 50%)
Stored energy for the Max. shock 42 J
Pacing amplitude during ATP
therapies
7 V (Actual value at 300 ms: 5.3 V )
Anti-PMT protection Termin
(1) The device has 50% tilt in each phase thus delivers 94% of stored energy. Each phase is limited to
10 ms duration.
66
SORIN PLATINIUM CRT-D U459A