User's Manual

Table Of Contents
14. IMPLANT PROCEDURE
NECESSARY EQUIPMENT
Implantation of PLATINIUM 4LV SonR CRT-D 1844 requires the following equipment:
Sorin ORCHESTRA programmer, equipped with the SMARTVIEW software interface
and inductive telemetry head,
Sorin ORCHESTRA PLUS programmer, equipped with the SMARTVIEW software
interface, inductive telemetry head and optionally ORCHESTRA PLUS LINK,
pacing system analyzer, as well as its sterile connecting cables, to evaluate the pacing
and sensing thresholds,
a complete set of leads with corresponding introducers,
physiological signal monitor capable of displaying simultaneously the surface ECG and
arterial pressure,
an external defibrillator with sterile external paddles,
sterile cover for the telemetry head.
NOTE: In case you are implanting a DF4 and/or IS4 lead(s), please verify its compatibility
with standard alligators pin; please refer to the lead user’s manual for more details.
PACKAGING
Contents
PLATINIUM 4LV SonR CRT-D 1844 and its accessories are ethylene oxide sterilized and
hermetically sealed in two-ply clear packaging meeting international requirements.
The sterile packaging contains:
the defibrillator
a ratcheting screwdriver
As delivered, PLATINIUM 4LV SonR CRT-D 1844 is programmed to as-shipped values that
are different from nominal values (see Chapter “Programmable Parameters” for details).
OPTIONAL EQUIPMENT
The following equipment may be required during implantation of PLATINIUM 4LV SonR CRT-
D 1844:
an IS-1 insulating plug to close the atrial port
sterile water to clean traces of blood. Any parts cleaned with sterile water must be
thoroughly dried.
mineral oil to lubricate if necessary
a lead cap to isolate a lead which is not used
14.1.
14.2.
14.2.1.
14.3.
14. IMPLANT PROCEDURE
SORIN – PLATINIUM 4LV SonR CRT-D 1844 – U906A 33