User's Manual
Table Of Contents
- 1. General description
- 2. Indications
- 3. Contraindications
- 4. Warnings and precautions
- 5. Adverse events
- 6. Clinical studies
- 7. Patient selection and treatment
- 8. Patient counselling information
- 9. Declaration of conformity
- 10. Physician guidelines
- 11. Patient information
- 12. How supplied
- 13. Device description
- 14. Implant procedure
- 14.1. Necessary equipment
- 14.2. Packaging
- 14.3. Optional equipment
- 14.4. Before opening the package
- 14.5. Prior to implantation
- 14.6. Device placement
- 14.7. Choosing the type of lead
- 14.8. Shock configuration (+ -> -)
- 14.9. Measurement of thresholds at implant
- 14.10. Leads connection
- 14.11. Device implantation
- 14.12. Tests and programming
- 15. Special modes
- 16. Main functions
- 17. Patient follow-up
- 18. Supplemental Information
- 19. Physical characteristics
- 20. Electrical characteristics
- 21. Programmable parameters
- 22. Non programmable parameters
- 23. Limited warranty
- 24. Patents
- 25. Explanation of symbols
14. IMPLANT PROCEDURE
NECESSARY EQUIPMENT
Implantation of PLATINIUM 4LV SonR CRT-D 1844 requires the following equipment:
― Sorin ORCHESTRA programmer, equipped with the SMARTVIEW software interface
and inductive telemetry head,
― Sorin ORCHESTRA PLUS programmer, equipped with the SMARTVIEW software
interface, inductive telemetry head and optionally ORCHESTRA PLUS LINK,
― pacing system analyzer, as well as its sterile connecting cables, to evaluate the pacing
and sensing thresholds,
― a complete set of leads with corresponding introducers,
― physiological signal monitor capable of displaying simultaneously the surface ECG and
arterial pressure,
― an external defibrillator with sterile external paddles,
― sterile cover for the telemetry head.
NOTE: In case you are implanting a DF4 and/or IS4 lead(s), please verify its compatibility
with standard alligators pin; please refer to the lead user’s manual for more details.
PACKAGING
Contents
PLATINIUM 4LV SonR CRT-D 1844 and its accessories are ethylene oxide sterilized and
hermetically sealed in two-ply clear packaging meeting international requirements.
The sterile packaging contains:
― the defibrillator
― a ratcheting screwdriver
As delivered, PLATINIUM 4LV SonR CRT-D 1844 is programmed to as-shipped values that
are different from nominal values (see Chapter “Programmable Parameters” for details).
OPTIONAL EQUIPMENT
The following equipment may be required during implantation of PLATINIUM 4LV SonR CRT-
D 1844:
― an IS-1 insulating plug to close the atrial port
― sterile water to clean traces of blood. Any parts cleaned with sterile water must be
thoroughly dried.
― mineral oil to lubricate if necessary
― a lead cap to isolate a lead which is not used
14.1.
14.2.
14.2.1.
14.3.
14. IMPLANT PROCEDURE
SORIN – PLATINIUM 4LV SonR CRT-D 1844 – U906A 33