VENTRICULAR TACHYARRHYTHMIA THERAPIES
Common parame-
ters
Values Nominal value "As shipped"
value
Enable ATP therapy
Yes-No No No
Enable shock thera-
py
Yes-No Yes No
ATP pacing chamber
Right-R+L-L+R Right Right
Polarity alternation
(42J)
Yes-No No No
Atrial coil (SVC) pre-
sent
Yes-No Yes
Active case
Yes-No Yes
Shock configuration
(+ --> -)
Case to RV-SVC to RV-
Case + SVC to RV-RV to Case-
RV to SVC-RV to Case + SVC
RV to Case +
SVC
RV to Case +
SVC
SVC exclusion
(shock < 15J)
Yes-No No No
Therapy parameters in slow VT zone
ATP 1 program Values Nominal value "As shipped"
value
ATP program
OFF-Burst-Burst+Scan-Ramp-
Ramp+Scan
OFF OFF
Number of se-
quences
1-2-3-4-5-6-7-8-9-10-11-12-13-14-15 3 3
Cycles in first se-
quence
1-2-3-4-5-6-7-8-9-10-11-12-13-14-15 8 8
Cycles added per
sequence
0-1-2-3-4-5-6-7-8-9-10-11-12-13-14-15 0 0
Coupling interval (%)
50-55-60-65-70-75-80-85-90-95 80 80
Ramp decrement
(per cycle) (ms)
0-4-8-12-16-20-30-40-50-60 0 0
Scan decrement (per
sequence) (ms)
0-4-8-12-16-20-30-40-50-60 8 8
Time limit (min)
0.5-1-1.5-2-2.5-3-3.5-4 2 2
Minimum cycle
length (ms)
95-110-125-140-155-170-190-205-220-
235-250-265-280-295-310
220 220
21.3.
21.3.1.
21. PROGRAMMABLE PARAMETERS
SORIN – PLATINIUM 4LV SonR CRT-D 1844 – U906A 63