User's Manual

Table Of Contents
Shock program Values Nominal value "As shipped"
value
Shock 1 (J)
OFF-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-6-
7-8-9
10-12-14-16-18-20-22-24-26-28-30-32-
34-42
OFF OFF
Shock 2 (J)
OFF-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-6-
7-8-9
10-12-14-16-18-20-22-24-26-28-30-32-
34-42
OFF OFF
Number of Max.
Shock (42 J)
OFF-1-2-3-4 4 4
Therapy parameters in fast VT / VF zone
ATP 1 program Values Nominal value "As shipped"
value
ATP program
OFF-Burst-Burst+Scan-Ramp-
Ramp+Scan
Burst Burst
Number of se-
quences
1-2-3-4-5-6-7-8-9-10-11-12-13-14-15 1 1
Cycles in first se-
quence
1-2-3-4-5-6-7-8-9-10-11-12-13-14-15 8 8
Cycles added per
sequence
0-1-2-3-4-5-6-7-8-9-10-11-12-13-14-15 0 0
Coupling interval (%)
50-55-60-65-70-75-80-85-90-95 80 80
Ramp decrement
(per cycle) (ms)
0-4-8-12-16-20-30-40-50-60 0 0
Scan decrement (per
sequence) (ms)
0-4-8-12-16-20-30-40-50-60 0 0
Time limit
10s-20s-30s-1min-1.5min-2min 30s 30s
Minimum cycle
length (ms)
95-110-125-140-155-170-190-205-220-
235-250-265-280-295-310
205 205
Shock program Values Nominal value "As shipped"
value
Shock 1 (J)
OFF-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-6-
7-8-9
10-12-14-16-18-20-22-24-26-28-30-32-
34-42
OFF OFF
Shock 2 (J)
OFF-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-6-
7-8-9
10-12-14-16-18-20-22-24-26-28-30-32-
34-42
OFF OFF
Number of Max.
Shock (42 J)
1-2-3-4 4 4
REMOTE ALERTS AND WARNINGS
The device routinely performs security self-checks and technical measurements to ensure
system integrity. When system integrity is found to be at risk outside a follow-up, alerts are
stored in the device memory. When system integrity is found to be at risk during a follow-
21.3.3.
21.4.
21. PROGRAMMABLE PARAMETERS
66 SORIN – PLATINIUM 4LV SonR CRT-D 1844 – U906A