Thirty-five patients contributed data to evaluate the percentage of
ventricular pacing provided with SafeR. Twenty-nine patients had 1
% or less ventricular pacing and six patients had a range of 28-97 %
ventricular pacing. The graph below shows the distribution of
ventricular pacing observed in patients with and without AV block as
a primary indication for implant.
The graph shows that many patients programmed to SafeR had less
than 1% ventricular pacing:
84 % of patients without AV block at implant.
63 % of patients with AV block at implant.