User's Manual

48 US-ENGLISH
15.4. DETECTION CHARACTERISTICS IN THE
PRESENCE OF ELECTROMAGNETIC FIELDS
Per Clause 27.4 of Standard EN 45502-2-2, detection in the
presence of electromagnetic fields is characterized as follows:
Differential mode:
Common mode rejection ratio:
16.6 Hz
50 Hz
60 Hz
Atrial channel
≥ 75 dB
67 dB
67 dB
Ventricular channel
≥ 69 dB
69 dB
≥ 69 dB
Modulated interference: For atrial sensitivity setting of 0.2 mV,
compliance to the Cenelec standard 45502-2-2 is met for a maximum
test signal amplitude of 8 V for the frequency of 60 MHz. 0.4 mV
complies with the standard for the whole frequency range.