17. PATIENT FOLLOW-UP
17.1. FOLLOW-UP RECOMMENDATIONS
Before the patient is discharged and at each subsequent follow-up
visit, it is advisable to:
check the occurrence of system warnings
check the battery status,
check the integrity of the pacing and defibrillation leads,
check for proper sensing (sensitivity, crosstalk) and
pacing ; set the pacing amplitude to twice the pacing
threshold,
interrogate the implant memories (AIDA+),
check the efficacy of the therapies delivered,
keep a printout of programmed parameters, test results,
and memory data,
reset the memory data and statistics.
These operations should be performed by medical personnel in an
appropriate care unit, with resuscitation equipment present.
It is recommended that a routine follow-up examination be done one
month after discharge, and then every three months until the device
nears the replacement date.
After a device reset, the magnet rate is equal to 87 ppm; it will be
updated within the next 24 hours.
Refer to the online help for a description of displayed warning, and the
necessity to contact Sorin CRM for an evaluation.
Implant software upgrade: In case a new implant software is
downloaded in the device memory through the programmer, a
warning message could be displayed by the programmer to inform
the user and give the proper instructions to follow.