17.2. HOLTER FUNCTION
The Holter records up to 14 tachyarrhythmia episodes as well as the
therapy history.
STORED EPISODES
PARADYM RF DR 9550 stores up to 14 episodes (VF, VT, Slow VT,
SVT/ST, nonsustained).
For each episode four levels of details are presented:
Tachogram (to visualize PP and PR intervals)
Event log for the entire episode:
PARAD/PARAD+ analysis for each majority,
Delivered therapies,
Markers: Atrial and ventricular markers, sensed, paced and in
relative refractory periods,
EGM: onset and detection of the arrhythmia, on two therapies,
and the return to slow rhythm by recording electrogram.
Therapy history: For each arrhythmia detection, each therapy
delivered (either automatically or during an electrophysiological
study) and at the end of each arrhythmia, PARADYM RF DR 9550
records the type of majority rhythm, the number of ATP sequences
delivered, the energy and the number of shocks delivered.