User's Manual

68 US-ENGLISH
Pacing/Sensing
Values
Atrial sensitivity (mV)
(1)
From 0.2 to 4 by steps of 0.2 ; 0.4
50 %)
Atrial amplitude (V)
(2)
1-1.5-2-2.5-3-3.5-4-4.5-5-6 (± 20 %)
Atrial pulse width (ms)
0.12-0.25-0.35-0.5-0.6-0.75-0.85-1
10 %)
Ventricular sensitivity (mV)
(1)
From 0.4 to 4 by steps of 0.2 ; 0.4
50 %)
Ventricular amplitude (V)
(2)
1-1.5-2-2.5-3-3.5-4-4.5-5-6 (± 20 %)
Ventricular pulse width (ms)
0.12-0.25-0.35-0.5-0.6-0.75-0.85-1
10 %)
(1) Values are measured using a positive and negative triangular signal of
2/13 ms.
(2) The correlation between the programmed amplitudes, the stored
amplitudes and the mid-pulse delivered amplitudes under a 500 ohm
load are given in the following table:
Programmed ampl. (V)
1
1.5
2
2.5
3
3.5
Mid-pulse delivered ampl. (V)
0.97
1.39
1.79
2.35
2.84
3.25
Stored amplitude (V)
1.14
1.63
2.1
2.76
3.33
3.82
Programmed ampl. (V)
4
4.5
5
6
Mid-pulse delivered ampl. (V)
3.58
4.23
4.47
5.37
Stored amplitude (V)
4.2
4.96
5.25
6.3