US-ENGLISH 73
20.3. VENTRICULAR TACHYARRHYTHMIA
THERAPIES
Common parameters
Values
Enable ATP therapy
Yes-No
Enable shock therapy
Yes-No
Polarity alternation (42J)
Yes-No
Atrial coil (SVC) present
Yes-No
Active case
Yes-No
Shock configuration (+ --> -)
Case to RV-SVC to RV-Case + SVC to
RV-RV to Case-RV to SVC-RV to
Case + SVC
SVC exclusion (shock < 15J)
Yes-No
Therapy parameters in slow VT zone
ATP 1 program
Values
ATP program
OFF-Burst-Burst+Scan-Ramp-
Ramp+Scan
Number of sequences
1-2-3-4-5-6-7-8-9-10-11-12-13-14-15
Cycles in first sequence
1-2-3-4-5-6-7-8-9-10-11-12-13-14-15
Cycles added per sequence
0-1-2-3-4-5-6-7-8-9-10-11-12-13-14-15
Coupling interval (%)
50-55-60-65-70-75-80-85-90-95
Ramp decrement (per cycle) (ms)
0-4-8-12-16-20-30-40-50-60
Scan decrement (per sequence) (ms)
0-4-8-12-16-20-30-40-50-60
Time limit (min)
0.5-1-1.5-2-2.5-3-3.5-4
Minimum cycle length (ms)
95-110-125-140-155-170-190-205-220-
235-250-265-280-295-310