User's Manual

Power supply AC input: 100-240V ~ 50-60Hz 600
mA
(to wall socket) Class II
Telemetry
head
Type BF applied part on the lower
side of telemetry head (applied part
temperature can reach a maximum of
45°C/113°F)
Operating
conditions
from 0°C to +35°C (32 °F to 95 °F)
from 5% to 93% RH non condensing
Transport and
storage
conditions
from -15°C to +70°C (5 °F to 158 °F)
from 5% to 90% RH non condensing
Transport and
storage
Atmospheric
air-pressure
limits
Min: 700 hPa
Max: 1060 hPa
Telemetry
head
transmitter
frequency ~25 kHz modulation pulsed
-10 dBμA/m at 10m
Telemetry
head receiver
Preferred frequencies: 8 KHz, 16 KHz
Modem
Class 4 (33dBm ± 2dB) for GSM 850
Class 1 (33dBm ± 2dB) for GSM
1900
Class 3 (24dBm +1/-3dB) for
WCDMA (B2/B4/B5)
Class 3 (23dBm ± 2dB) for LTE FDD
(B2/B4/B5/B7/B12/B13/B25/B26)
Class 3 (23dBm ± 2dB) for LTE TDD
(B38/B41)
FCC approvals
11.1. RADIO EQUIPMENT EMISSION
Radio
Equipment
Transmitter
Frequency
Bands/ Maximal
Power
Receiver
Frequency
Bands
Inductive
Telemetry
CPR4BR
frequency ~25 kHz
modulation pulsed
-10 dBµA/m at 10m
8kHz, 16kHz
Radio
Equipment
Transmitter
Frequency
Bands/ Maximal
Power
Receiver
Frequency
Bands
Modem
2G/3G/4G
Quectel
EG25-G
Class 4 (33dBm ±
2dB)
for GSM 850
Class 1 (33dBm ±
2dB)
for GSM 1900
Class 3 (24dBm
+1/-3dB)
for WCDMA
(B2/B4/B5)
Class 3 (23dBm ±
2dB)
for LTE FDD
(B2/B4/B5/B7/
B12/B13/B25/B26)
Class 3 (23dBm ±
2dB)
for LTE TDD
(B38/B41)
FCC approvals
B2: 1900 MHz
B4: 1700 MHz
B5: 850 MHz
B7: 2600 MHz
B12: 700 MHz
B13: 700 MHz
B25: 1900
MHz
B26: 850 MHz
B38: 2600
MHz
B41: 2500
MHz
The Smart Spot may be subject to
interference by other equipment, even if the
equipment complies with CISPR emission
requirements.
11.2. RAW MATERIALS
The following materials (CPR4BR) may remain in
contact with body tissues.
Item
Material
Description
Plastic
parts
Cycoloy C2100
HF (ABS/PC,
UL 94 V-0,
White
RAL9016)
POLYCARBONATE ABS
(ABS-PC) White for outer
parts and grey for the
middle part.
Label
Glossy white
polyesther B423
POLYESTER
Main
cable
Polyurethane
(UL 94 V-2)
POLYURETHANE
12. DATA PROTECTION
MicroPort CRM undertakes to take all necessary security measures to protect the clinical data and in
particular to strictly restrict access to it. For any information regarding processing of your personal data,
please refer to the information and patient consent that you have signed with your physician who is the data
controller.
13. DECLARATION OF CONFORMITY
The FCC Product ID is YSGKC861, which contains FCC ID: XMR201903EG25G.
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant
to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful
interference in a residential installation. This equipment generates uses and can radiate radio frequency
energy and, if not installed and used in accordance with the instructions, may cause harmful interference
to radio communications. However, there is no guarantee that interference will not occur in a particular
installation.
Smart Spot – UA10275A 7