17. PATIENT FOLLOW-UP
17. PATIENT FOLLOW-UP
17.1. FOLLOW-UP RECOMMENDATIONS
Before the patient is discharged and at each subsequent follow-up visit, it is advisable to:
check the occurrence of system warnings
check the battery status,
check the integrity of the pacing and defibrillation leads,
check for proper sensing (sensitivity, crosstalk) and pacing ; set the pacing amplitude to
twice the pacing threshold,
interrogate the implant memories (AIDA+),
check the efficacy of the therapies delivered,
keep a printout of programmed parameters, test results, and memory data,
reset the memory data and statistics.
These operations should be performed by medical personnel in an appropriate care unit,
with resuscitation equipment present.
It is recommended that a routine follow-up examination be done one month after discharge,
and then every three months until the device nears the replacement date.
After a device reset, the magnet rate is equal to 87 bpm; it will be updated within the next 24
hours.
Refer to the online help for a description of displayed warning, and the necessity to contact
Sorin for an evaluation.
Implant software upgrade:
In case a new implant software is downloaded in the device memory through the
programmer, a warning message could be displayed by the programmer to inform the user
and give the proper instructions to follow.
17.2. HOLTER FUNCTION
The Holter records up to 14 tachyarrhythmia episodes as well as the therapy history.
Stored Episodes
INTENSIA SonR CRT-D 184 stores up to 14 episodes (VF, VT, Slow VT, SVT/ST, non-
sustained).
For each episode four levels of details are presented:
Tachogram (to visualize PP and PR intervals)
Event log for the entire episode:
PARAD/PARAD+ analysis for each majority,
Delivered therapies,
SORIN INTENSIA SonR CRT-D 184 U150A
43
NOTE: If the last reforming, charge or shock occurred during the week preceding the
interrogation, the last battery value may be still impacted by the event. One week post
event, the battery will recover its steady state value.