User Manual
Table Of Contents
- 1. General description
- 2. Indications
- 3. Contraindications
- 4. Warnings and precautions
- 5. Adverse events
- 6. Clinical studies
- 7. Patient selection and treatment
- 8. Patient counselling information
- 9. Conformance to standards
- 10. Physician guidelines
- 11. Patient information
- 12. How supplied
- 13. Device description
- 14. Implant procedure
- 14.1. Necessary equipment
- 14.2. Packaging
- 14.3. Optional equipment
- 14.4. Before opening the package
- 14.5. Prior to implantation
- 14.6. Device placement
- 14.7. Choosing the type of lead
- 14.8. Shock configuration (+ -> -)
- 14.9. Measurement of thresholds at implant
- 14.10. Leads connection
- 14.11. Device implantation
- 14.12. Tests and programming
- 15. Special modes
- 16. Main functions
- 17. Patient follow-up
- 18. Supplemental Information
- 19. Physical characteristics
- 20. Electrical characteristics
- 21. Programmable parameters
- 22. Non programmable parameters
- 23. Limited warranty
- 24. Patents
- 25. Explanation of symbols
20. ELECTRICAL CHARACTERISTICS
Total capacity 1964 mAh
Usable capacity Between BOS and RRT: 1278 mAh. Between BOS and EOS:
1675 mAh.
Voltage BOS: 3.25 V. RRT: 2.66 V. EOS: 2.5 V.
20.3. LONGEVITY
The longevities mentioned below are calculated by taking into account 6 months storage.
5.1 years Biventricular pacing in DDD mode, 100%, 500 ohm, 3.5 V, 0.35 ms, 60 bpm,
one 42 J shock per quarter, sensors OFF
4.6 years Biventricular pacing in DDD mode, 100%, 500 ohm, 3.5 V, 0.35 ms, 60 bpm,
one 42 J shock per quarter, sensors (G, SonR) ON
6.0 years Biventricular pacing in DDD mode, 1% in atrium, 100% in both ventricles, 500
ohm, 3.5 V, 0.35 ms, 60 bpm, one 42 J shock per quarter, sensors OFF
4.2 years Biventricular pacing in DDD mode, 15% in atrium, 100% in both ventricles,
500 ohm, 4.5 V, 0.50 ms, 60 bpm, one 42 J shock per quarter, sensors OFF
9.0 years 0% pacing, one 42 J shock per quarter, sensors OFF
The mean longevity as a function of shocks delivered at maximum energy, with and without
pacing, is as follows:
The mean longevity as a function of yearly remote follow-ups
(1)
, with and without pacing, is
as follows:
(1) An excessive number of remote follow-ups can have a non-negligible impact on device
longevity.