21. PROGRAMMABLE PARAMETERS
Ventricular arrhythmia algorithms Values
Atrial pacing on PVC Yes-No
Post extrasystolic pause suppression Yes-No
Acceleration on PVC Yes-No
Max acceleration rate (bpm) From 60 to 145 by steps of 5 ; 100
Post-shock mode Values
Mode OFF-VVI-DDI-DDD
Duration 10s-20s-30s-1min-2min-3min-4min-5min
Basic rate (bpm) From 50 to 90 by steps of 5 ; 60 (± 4 %)
Rest AV delay (ms) 30-40-45-55-65-70-80-85-95-100-110-115-125-135-140-150-
155-165-170-180-190-195-205-210-220-225-235-250
(± 19 ms)
Exercise AV delay (ms) 30-40-45-55-65-70-80-85-95-100-110-115-125-135-140-150-
155-165-170-180-190-195-205-210-220-225-235-250
(± 19 ms)
AVD Paced/Sensed Offset (ms) 0-10-15-25-30-40-45-55-65-70-80-85-95-100-110-115-125
(± 1 ms)
A amplitude (V) 1-1.5-2-2.5-3-3.5-4-4.5-5-6 (± 20 %)
A pulse width (ms) 0.12-0.25-0.35-0.5-0.6-0.75-0.85-1 (± 10 %)
RV amplitude (V) 1-1.5-2-2.5-3-3.5-4-4.5-5-6 (± 20 %)
RV pulse width (ms) 0.12-0.25-0.35-0.5-0.6-0.75-0.85-1 (± 10 %)
LV amplitude (V) 0.25- (± 50 %)
0.5-0.75- (± 30 %)
1-1.25-1.5-1.75-2-2.25-2.5-2.75-3-3.25-3.5-3.75-4-4.25-4.5-
4.75-5-6-7 (± 20 %)
LV pulse width (ms) 0.12-0.25-0.35-0.5-0.6-0.75-0.85-1 (± 10 %)
Refractory periods Values
Atrial refractory period post
ventricular sensing (ms)
45-65-80-95-110-125-140-155 (± 16 ms)
Atrial refractory period post
ventricular pacing (ms)
80-95-110-125-140-155 (± 4 ms)
Sensitivity margins Values
Atrial post pacing/sensing margin
(mV)
From 0 to 1 by steps of 0.2 ; 0.4
Ventricular post pacing margin (mV) From 0 to 2 by steps of 0.2 ; 0.8