User Manual

Table Of Contents
21. PROGRAMMABLE PARAMETERS
Response to noise Values
Automatic sensitivity on noise ON-OFF
V pacing on noise ON-OFF
SafeR (AAI <=> DDD) parameters Values
AVB I switch Rest+Exercise-Exercise
Long PR: max (ms) From 200 to 500 by steps of 50 ; 450
Long PR: min (ms) From 200 to 500 by steps of 50 ; 250
Max. pause (s) 2-3-4
21.2. VENTRICULAR TACHYARRHYTHMIA DETECTION
Therapy zones Values
Slow VT detection zone
(1)
Slow VT ON-Slow VT OFF
VT detection zone VT ON-VT OFF
Fast VT / VF detection zone Fast VT+VF ON-VF ON
Slow VT rate (lower limit) (bpm) From 100 to 200 by steps of 5 ; 190
VT rate (lower limit) (bpm) 130-135-140-145-150-155-160-165-170-175-180-185-190-
195-200-210-220-230
VF rate (lower limit) (bpm) 150-155-160-165-170-175-180-185-190-195-200-210-220-
230-240
Fast VT rate (upper limit) (bpm) 155-160-165-170-175-180-185-190-195-200-210-220-230-
240-255
Slow VT persistence (cycles) 4-6-8-12-16-20-30-50-100-200
VT persistence (cycles) 4-6-8-12-16-20-30-50-100-200
VF persistence (cycles) From 4 to 20 by steps of 1 ; 6
(1) The Slow VT zone should be programmed ON only if the VT zone is programmed ON.
SORIN INTENSIA SonR CRT-D 184 U150A
55