14.7. CHOOSING THE TYPE OF LEAD
The defibrillator should be connected to:
one ventricular defibrillation lead with sensing/pacing bipolar
electrodes, and one or two defibrillation electrodes.
The choice of leads and their configuration is left to the implanting
physician’s judgment.
Connectors: The bipolar pacing/sensing connector is compliant with
the IS-1 standard and the defibrillation connectors are compliant with
the DF-1 standard.
Shock configuration (+ -> -)
The shock configuration is the energy pathway between the
defibrillation electrodes. If an atrial coil is present, the shock
configuration can be programmed for bi-directional shocks.
Programming: When active case and SVC are both programmed to
Yes, the shock configuration can be programmed to: 1. RV to Case
(or Case to RV), 2. RV to SVC (or SVC to RV), 3. RV to Case+SVC
(or Case+SVC to RV).
The polarity of shock is determined by the parameter itself.
RV to Case+SVC
RV to Case
RV to SVC