User's Manual

46 US-ENGLISH
User website
On the website, the physician is able to:
consult and schedule the remote follow-ups of their patient
configure additional ways of being notified of alerts (for instance
by SMS, fax or e-mail)
consult, print and export patient reports
Alert system
The following set of alert trigger can be independently programmed
ON/OFF by the physician using the Orchestra Plus programmer and
can trigger an alert transmission:
Reset of the device
ERI reached
Low or high impedance (A, RV, LV)
Abnormal coil impedance (shock lead)
Low or High shock impedance
Long charge time
Inefficient high energy shock
All shocks programmed OFF
Shock treated VT/VF
Lack of V pacing in CRT device
Suspicion of noise on the V lead
Fast V rate during AF
WARNINGS
The use of remote monitoring does not replace regular follow-up.
Therefore, when using remote monitoring, the time period between
follow-ups visits may not be extended.
When ERI mode is reached, this information is transmitted via the
remote monitoring facility and then the remote-monitoring is switched
off to preserve battery life.