User's Manual

US-ENGLISH 61
Detection criteria
Values
Slow VT and VT detection criteria
Rate Only-Stability-Stability+-
Stability/Acc-Stability+/Acc
Fast VT detection criteria
Rate+Stability-Rate Only
Majority: (X/Y), Y (cycles)
8-12-16
Majority: (X/Y), X (%)
65-70-75-80-90-95-100
Window of RR stability for Slow VT
and VT (ms)
30-45-65-80-95-110-125-125
Window of RR stability for fast VT (ms)
30-45-65
Prematurity acceleration (%)
6-13-19-25-31-38-44-50
Long cycle persistence extension
(cycles)
From 0 to 16 by steps of 1; 10
Long cycle gap (ms)
15-30-45-65-80-95-110-125-140-155-
170-190-205
20.3. VENTRICULAR TACHYARRHYTHMIA
THERAPIES
Common parameters
Values
Enable ATP therapy
Yes-No
Enable shock therapy
Yes-No
Polarity alternation (42J)
Yes-No
Atrial coil (SVC) present
Yes-No
Active case
Yes-No
Shock configuration (+ --> -)
Case to RV-SVC to RV-Case + SVC to RV-
RV to Case-RV to SVC-RV to Case + SVC
SVC exclusion (shock < 15J)
Yes-No