User's Manual

www.spacelabshealthcare.com 061-2050-00 Rev. D 9/09
91367 SL2200 Compact Monitor
Z Wireless Ethernet
Communication; provides
802.11b wireless ethernet with
WEP security; requires site
qualification survey to ensure
adequate performance
Product Specifications
Classification
MDD Class IIb
EN 60601-1
Class 1 Requires outlet with
safety ground (Protective Earth)
conductor. Rated for continuous
operation.
CISPR11, Group 1, Class B
Suitable for use in domestic
establishments connected to a
low-voltage supply network
Electrical Specifications
Mains Power 100 to 240 VAC,
50 to 60 Hz, 2 A
Batteries One or two 12-volt
(2.45 Ah NiMH) batteries may
be used, providing 2.5 to
4 hours of operation with two
batteries (TFT display),
depending on monitor
configuration and setup, for up
to 600 charge/discharge cycles;
operating time is dependent on
configuration and usage;
2 hours are required to charge
batteries to 100% of capacity
with AC connected and the
monitor powered OFF; 3-hour
charge time required with
monitor powered ON.
Isolation Chassis leakage
current meets AAMI,
UL 60601-1, CSA Std. C22.2
No. 601.1, and IEC 60601-1
standards
Physical Dimensions
Height
27.05 cm (10.65 inches)
Depth
16.51 cm (6.5 inches)
Width
31.24 cm (12.3 inches)
Weight
5.5 kg (12.1 pounds) with
two batteries and recorder
Environmental Requirements
Storage
Temperature
-25° to 60° C (-13° to 140° F)
Humidity
95% (non-condensing)
Altitude
0 to 12,192 meters
(0 to 40,000 feet)
Operating
Temperature
0° to 50° C (32° to 122° F)
Humidity
95% (non-condensing)
Altitude
0 to 4,572 meters
(0 to 15,000 feet)
Electromagnetic Compatibility
EN 60601-1-2: 2001
Emissions (CISPR 11)
EN 55011, Class B
EN 61000-3-2
Harmonics
EN 61000-3-3
Flicker
Immunity
IEC 61000-4-2
ESD, 8 kV contact/15 kV air
IEC 61000-4-3
RF fields, 20 volts/meter,
80 MHz to 2.5 GHz
IEC 61000-4-4
Burst, 1 kV data and
I/O ports/2 kV power ports
IEC 61000-4-5
Surge, 1 kV differential/
2 kV common mode
IEC 61000-4-6
Conducted RF, 3 Vrms
150 kHz to 80 MHz