TELEMETRY TRANSMITTER (96281) OPERATIONS MANUAL 3-18
AMBIENT LIGHT INTERFERENCE — Cover sensor area
The sensor is receiving external light interference from a bright light source near the sensor. Shield
the sensor from the external light source.
The sensor photodiode and LED are misaligned on flexible sensors, allowing light to enter. Realign
the sensor photodiode with the LED.
If the message persists, then replace the sensor.
NOISY SIGNAL
The sensor signal is disturbed by motion or other interference. Eliminate sensor movement.
Power cords or other electrically noisy devices are too close to the sensor. Move the noisy device
or move the sensor to another site.
If the message persists, then replace the sensor.
FAULTY SENSOR — Replace sensor
The LED or photodiode (or both) may have failed.
Ensure the sensor is properly connected, disconnect and then reconnect the sensor.
If the error occurs again, replace the sensor and/or sensor adapter cable.
If replacing the sensor and/or cable does not correct the problem, contact your Spacelabs
Healthcare Field Service Engineer.
HARDWARE INCOMPATIBILITY — Contact service
The hardware configuration in your module is not compatible with the programmed software options.
Contact your Spacelabs Healthcare Field Service Engineer.
Sensors
For SpO
2
sensor compatibility, refer to the information that follows.
Spacelabs Healthcare SpO
2
Sensors
Please refer to the the accessories in Telemetry Transmitter (96281) Data Sheet for compatible
sensors (for the 96281-C) and adapter cables.
Additional Information
For additional information about biocompatibility or sensor disposal, refer to the manufacturer's
instructions enclosed with each sensor.