User's Manual

Contents
7-5
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Display Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Setting User-Defined Default Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Storing Changed Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Selecting Alarm Attributes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Parameter Configurations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Transferring User-Defined Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Retrieving User-Defined Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Restoring Factory Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Restoring User-Defined Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Module Configuration Manager
Overview
The Module Configuration Manager feature provides the capability to define and
store all user-configurable options within your module. Once a module has been
configured, these settings will control its operation whenever the module is
powered ON.
To ensure standardization of multiple modules, settings may be transferred from
one configured module to another. Settings are first transferred from a configured
module to the monitor (using the TRANSFER SETTINGS key), then retrieved
from the monitor into other modules (using the RETRIEVE SETTINGS key).
The Restore Settings feature changes all the module’s parameter user-
configurable options to the defaults previously stored as user settings.
Depending on the parameter configuration of the module, the Module
Configuration Manager feature is accessible through either the ECG or TEMP
parameter key. For module configurations without ECG, use the TEMP key. For
module configurations with ECG, use the ECG key. (Refer to Module
Configuration Manager on page 15-23 for more information.)
User-configurable parameters are available based on the
options in your module. For example, if your module does not
have the Arrhythmia Processing feature (Multiview
I),
parameters related to arrhythmia detection and alarms will not
be available.
User-defined settings which have been transferred to the
monitor’s memory are available for retrieval into other modules
until (1) the monitor is powered OFF or (2) a patient is admitted
or discharged with a purge of data.
!
The TEMP key will only be displayed if a temperature probe is